Przeglądanie stron 10
Resistance
Range Resolution Accuracy
200 0.1 ± (0.8% rdg + 5 digits)
2000 1
20K 10
200K 100
± (0.8% rdg + 2 digits)
20M 10K ± (1.0% rdg + 5 digits)
Open Circuit Voltage: Less than 3V on all ranges
Overload Protection: All ranges 500Vrms (AC or DC Current).
Temperature
Range Resolution Accuracy
-40°F to 302°F ± (1.0% rdg + 5 digits)
303°F to 1832°F
1 degree Fahrenheit
± (1.5% rdg + 15 digits)
Overload Protection: 500Vrms
Thermocouple: K-Type
Note: The provided thermocouple is only intended to measure
Temperatures up to 300 degrees F. Exceeding this
temperature may damage the probe.
Diode Test and Continuity Beeper
Displays relative
reading in mV,
not actual diode voltage
Test current about 1mA.
Open circuit voltage about 2.8V
Beeper sounds if
meter displays <70.
(Approx. 30)
Test current about 1mA.
Open circuit voltage about 2.8V
Transistor hFE test
hFE
Can measure NPN or PNP
transistor hFE. Range: 0-1000 ß
Base current about
10µA,Vce about 2.8V
11
Przeglądanie stron 10