Triplett 9007-A Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Urządzenia pomiarowe Triplett 9007-A. Triplett 9007-A User Manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1

1 84-909 1/15

Strona 2 - 1: INTRODUCTION

10 2.32 Use caution when measuring circuits containing capacitors. Capacitors can store dangerous or lethal levels of electricity, even when th

Strona 3

11 injury to the user. 2.36 If there is any doubt about the condition of the meter (i.e. safe vs unsafe), remove the meter from service and secure

Strona 4

12 low side of a power circuit first. Never clamp onto a hot wire first, (usually red, black, or blue in AC wiring circuits.) If you clamp onto a

Strona 5

13 the hazards present. For example, the meter may indicate a low voltage when high voltages are actually present. Failure to observe this precautio

Strona 6

14 Discharging these capacitors can be dangerous unless an approved method is used. Failure to observe this precaution may result in damage to the m

Strona 7

15 3: INTERNATIONAL SYMBOLS The following International Symbols may be used in this manual and on the case of the meter to identify, caution, or w

Strona 8

16 4: PRODUCT FEATURES 4.1 30 Measurement Ranges 4.2 Huge 1.4” tall High Contrast LCD display 4.3 White LCD Backlight allows viewing in poorly l

Strona 9

17 5: SPECIFICATIONS 5.1 Display: ...1.4” (35mm) high LCD 5.2 Display Resolution: ...2000 counts, 0000 to 2

Strona 10

18 5: SPECIFICATIONS CONT. 5.13 Insulation: ...Double Insulated (Protection Class II) 5.14 Pollution Degree ….2 5.15 Approvals:

Strona 11

19 DC Voltage Range Resolution Accuracy 200.0mV 0.1mV ± (0.5% rdg + 2 digit) 2.000V 1mV 20.00V 10mV 200.0V 100mV 600V 1V ± (0.8% rdg + 2 dgt) In

Strona 12

2 TABLE OF CONTENTS 1. Introduction 2. Safety Rules and Warnings 3. International Symbols 4. Product Features 5. Specifications 6. Front Panel 7. Me

Strona 13

20 DC Current Range Resolution Accuracy 2.000mA 1µA ±(1.0% rdg + 3 digits) 200.0mA 100 µA ±(1.5% rdg + 3 digits) 10.00A 10mA ±(2.5% rdg + 10 digits

Strona 14

21 Resistance Range Resolution Accuracy 200.0Ω 0.1Ω ± (1.0% rdg + 4 digits) after subtracting any residual resistance noted when test leads ar

Strona 15

22 Sensor: Type K Thermocouple Overload protection: 250V DC or AC RMS. Diode Test and Continuity Be

Strona 16 - 4: PRODUCT FEATURES

23 6: FRONT PANEL . Large 2000 count Liquid Crystal Display . Power button: turns the meter ON or OFF. . Function Switch . 10A (positive) input j

Strona 17 - 5: SPECIFICATIONS

24 7: MEASUREMENT PROCEDURES WARNING! Do not use the meter when the low battery symbol is displayed. This may cause the meter to prod

Strona 18 - 5: SPECIFICATIONS CONT

25 7.2 Power Button and Auto Power Off: Turn on the 9007-A by pressing the yellow power button . Press it again to turn t

Strona 19 - Input

26 7.3.2 Set the RANGE switch to a V position. If the magnitude of the voltage is unknown, set the RANGE switch to the highest V position

Strona 20 -

27 measurement resolution can be obtained. 7.4 AC Voltage Measurement: WARNING! If the magnitude of the voltage to be measured is unknown, alway

Strona 21

28 7.4.4 If the display indicates overrange, i.e. “1- - -”, disconnect the test probes from the circuit and rotate the RANGE switch to the next h

Strona 22

29 7.5 DC Current Measurement: WARNING! If the magnitude of the current to be measured is unknown, always start by setting the meter to the highe

Strona 23 - 6: FRONT PANEL

3 2: SAFETY RULES & WARNINGS 2.1 Read all instructions in this manual before using this meter. Failure to do so may result in damage t

Strona 24 - 7: MEASUREMENT PROCEDURES

30 7.5.3 If the red test lead is inserted into the uA mA jack, set the RANGE switch to the 200m A position. Connect the test leads IN SERIES wi

Strona 25

31 7.6 AC Current Measurement: WARNING! If the magnitude of the current to be measured is unknown, always start by setting the meter to the highe

Strona 26

32 7.6.3 If the red test lead is inserted into the uA mA jack, set the RANGE switch to the 200m A~ position. Connect the test leads IN SE

Strona 27

33 7.7 Resistance Measurement: WARNING! Do not apply voltage or current to the meter when it is set to any of the “ Ω ” ranges. 7.7.1 Connect t

Strona 28

34 7.7.5 If the displayed value is less than “200” (decimal point not shown), a more accurate reading may be obtained by setting the RANGE switch

Strona 29

35 7.8 Continuity Beeper: WARNING! Do not apply voltage or current to the meter when it is set to the Diode Test / Continuity Beeper “

Strona 30

36 7.9 Diode Test: WARNING! Do not apply voltage or current to the meter when it is set to the Diode Test / Continuity Beeper “ ” range 7.9.1

Strona 31

37 7.10 Measuring Capacitance: WARNING! Do not apply voltage or current to the meter when it is set to the Capacitance “F ” ranges. Do not connect

Strona 32

38 7.10.5 If the displayed value is less than “200” (decimal point not shown), a more accurate reading may be obtained by setting the RANGE switch t

Strona 33

39 7.11 TEMPERATURE MEASUREMENTS: WARNING! To avoid electric shock, disconnect both test probes from any source of voltage before making a temperatu

Strona 34

4 2.5 Follow the recommendations of any Trade Organizations or Regulatory Agencies whose scope encompasses the use of this meter. Failure to do so

Strona 35

40 7.12.2 Insert the black test lead banana plug into the negative (-) jack (COM) and the red test lead banana plug into the positive (+) jack indic

Strona 36

41 9: MAINTENANCE Your Triplett Model 9007-A DMM is a precision measuring instrument and, when used as described in this manual, should not requ

Strona 37

42 9.1.3 Remove the 9 volt battery and, observing proper polarity, replace with a fresh battery (PN 37-48). 9.1.4 Reassemble battery compartment

Strona 38

43 9.2.5 Reassemble case of meter. Reinstall screws. 9.2.6 Verify that the meter operates properly before using it to make measurements.

Strona 39

44 10: ACCESSORIES 10.1 The Triplett Model 9007-A package contains the following items: - The Model 9007-A DMM - Test leads for the 9007-A (Tr

Strona 40

45 WARRANTY Triplett / Byte Brothers extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the orig

Strona 41 - 9: MAINTENANCE

46 ARE EXPRESSLY IN LIEU OF ANY OTHER WARRANTIES EXPRESSED OR IMPLIED. The purchaser agrees to assume all liability for any damages and bodily inju

Strona 44 - 10: ACCESSORIES

5 meter or injury to the user. 2.10 Make sure test leads are properly inserted and seated in the meter’s input jacks. A loose test lead may cause

Strona 45 - WARRANTY

6 the test leads in incorrect jacks can damage the meter and/or injure the user. 2.14 Do not exceed the maximum voltage or current limit

Strona 46

7 user. 2.19 Replace fuses only with exact or equivalent fuses. Do not “bridge” fuses out of circuit. Failure to observe this precaution may resu

Strona 47

8 this precaution may result in damage to the meter or injury to the user. 2.24 Do not use this meter to make measurements in adverse environments

Strona 48

9 this precaution may result in damage to the meter or injury to the user. 2.28 Do not modify the meter. Changing the design may make the meter

Komentarze do niniejszej Instrukcji

Brak uwag