6.1.8 Code Measurement Specifications
6.1.8.1 Peak Level Measurements
Peak Hold: Pointer will deflect to the peak level of
the input signal. For repetitive peaks
within a ±4% level window, 25 samples
are averaged, then no further updates
occur unless the level exceeds the
captured peak.
Peak Follow: Pointer will deflect to the peak level that
is at least 25% of the previously
displayed peak.
Level Stability: For a peak to be detected, the signal must
remain level stable within ±4.0% typically
for a period of 10 milliseconds.
AC Rejection for
DC Functions: AC signals greater than 60 Hz and up to
five times the level of the DC signal to be
measured will typically introduce less
than 3% additional error in the level
measurement.
DC Rejection for
AC Functions: DC signals up to five times the level of
the AC signal to be measured will
typically introduce less than 3%
additional error in the level measurement.